59
4.
Memory
- gives the operator the flexibility to change and edit the memory
location (1-50).
5.
Step (PROGRAM Mode only)
- gives the operator the flexibility to change
and edit the step location (1
– 9).
6.
Voltage
- gives the operator the flexibility to edit the voltage output.
7.
Frequency
- gives the operator the flexibility to edit the frequency output.
Frequency is only selectable when the source is configured for AC output.
8.
Transient (PROGRAM Mode only)
- gives the operator the flexibility to
enable or disable the transient function. Note: Transient Mode simulates the
effect of high speed voltage spikes and sags. For this reason, when
Transient Mode is set to ON, test parameters 9 through 28 will no longer be
displayed. Transient is only selectable when the source is configured for AC
output.
9.
Ramp Up (PROGRAM Mode only)
- gives the operator the flexibility to
increase the voltage output over a duration of time prior to achieving the
programmed output voltage.
10.
Timer Unit (PROGRAM Mode only)
- determines the time increment that
will be used for testing when the source is in PROGRAM Mode. The operator
can select between Second/Minute/Hour.
11.
Delay (PROGRAM Mode only)
- gives the operator the flexibility to
program a time delay, or warm up time. There is a voltage output present from
the source, but the high and low limit thresholds are essentially ignored during
this period.
12.
Dwell (PROGRAM Mode only)
- gives the operator the flexibility to
program the actual test time. This time begins after the ramp up time has
completed. The high and low limit thresholds are active once the delay time
has completed.
13.
Ramp Down (PROGRAM Mode only)
- gives the operator the flexibility to
program a time duration in which the output voltage is reduced to zero after
the dwell time has completed.
14.
Step Cycle (PROGRAM Mode only)
- gives the operator the flexibility to
program the number of looping cycles for a particular step. For example if the
operator would like to have a step repeat five times the step cycle would be
programmed to five. Step Cycle is only selectable when the source is
configured for AC output.