4.4 Connecting the power supply
Connect the mini-DIN-plug from the mains receiver to the port labelled
Pow
(“Power”) on
the lower right-hand side of your
iks aquastar midi
. Please pay attention to the marking on
the plug (upward pointing arrow). Now connect the mains receiver to the mains supply. It
must be located in a position that is
protected from water
(as generally applies for all
electrical connections). So the best location for the mains receiver is somewhere above
the water-line.
When putting the instrument into operation for the first time, you should perform a
Reset
(see Point 5.1).
4.5 Connection to the PC
With the serial interface lead that's available as an optional extra, the
iks aquastar
can be
connected to a PC for transferring the stored measurements. To do this, the mini-DIN plug
is connected to the mini-DIN- port
labelled PC
on
aquastar
. The 25-pin SUB-D plug is
connected to the PC's
serial interface
. The chain of measurements can then be
transferred to the PC with
iks datastar
software for further processing.
4.6 Connecting the external display
The optional external display is connected to the
iks aquastar's
mini-DIN-port labelled PC
with the mini-DIN-plug. Once the mains receiver, supplied and permanently connected to
the display, has been connected to the mains, the display will commence operation without
further adjustments.
The external display can be fitted either inset or mounted on a surface.
For inset installation a hole measuring 19.5 cm x 5 cm must be made at a suitable
location. Please note that the display must be fitted into the hole from
the inside to the
outside
(The mains receiver is too large to fit through the hole). The included short screws
are for securing the display.
For surface mounting, please use the long screws, plastic dowels and spacers.
5 Before you start
Please read the following section of the directions for use very carefully. Pay particular
attention to the areas where functions you want to use are described.
5.1 The reset function
As a general principal you should always carry out a so-called
Reset
. This erases
all
former programming
and returns the system to its
original condition
. Please proceed
as follows:
·
Disconnect the aquastar from the power supply
·
Now press the key labelled
del
and
keep it pressed.
·
Reconnect the power supply (continue to keep del pressed)
The following message appears on the display shortly afterwards:
·
Press the
yes
key again (or press
no
to abort and return to the main menu)
aquamidi V1.0
Reset yes/no