c
B
o
t
R
e
a
g
e
n
t
P
la
te
s
cBot 2 System Guide
9
cBot Reagent Plates
The configuration of the reagent plate differs between kit types, including the number of
rows that contain reagent.
Each 8-tube strip is labeled with the reagent name followed by a number. The number
indicates the row that it occupies on the reagent plate. If an 8-tube strip becomes dislodged,
use the row number on the label to return the tube strip to the appropriate position.
Flow Cell Type
Reagent Plate Description
HiSeq X and
HiSeq 3000/4000
Contains 12 rows with 8 deep wells each. Each reagent occupies a full
row of 8 wells. Not all rows contain reagent.
HiSeq High Output
(HiSeq v4)
Contains 8 rows of foil-sealed 8-tube strips filled with cluster
generation reagents. Rows 9 through 12 are empty.
HiSeq High Output
(TruSeq v3)
Contains 11 rows of foil-sealed 8-tube strips filled with cluster
generation reagents. Row 12 is empty.
HiSeq Rapid
Contains 3 rows of foil-sealed 8-tube strips filled with template
hybridization and first extension reagents. Rows 4 through 12 are
empty.
WARNING
Except for the HiSeq rapid reagent plate, these sets of reagents contain formamide, an
aliphatic amide that is a probable reproductive toxin. Personal injury can occur through
inhalation, ingestion, skin contact, and eye contact. Dispose of containers and any unused
contents in accordance with the governmental safety standards for your region.
For more
information, see the SDS for this kit at
.
Summary of Contents for cBot
Page 4: ...iv Material 20004364 Document 15065681 v01 ...
Page 16: ...10 Material 20004364 Document 15065681 v01 ...
Page 22: ...16 Material 20004364 Document 15065681 v01 ...
Page 66: ...60 Material 20004364 Document 15065681 v01 ...
Page 76: ...70 Material 20004364 Document 15065681 v01 ...
Page 88: ...Index 82 Material 20004364 Document 15065681 v01 ...