cBot 2 System Guide
65
Change the Adapter Plate
You can use a HiSeq flow cell or a GAIIx flow cell on the cBot. Each flow cell type requires
that a specific adapter plate is installed. Icons on the Start screen indicate which adapter
plate is installed.
NOTE
The cBot is shipped with the HiSeq adapter plate installed.
1
Open the instrument lid by gently lifting from the cutout on the front of the lid.
2
Lift the flow cell clamp.
3
Loosen the 2 captive Phillips head screws securing the adapter plate.
Figure 29
Flow Cell Adapter Plate
A
Captive screws
B
Adapter plate
4
Lift the existing adapter plate from the thermal stage and set aside.
5
If salts are present on the thermal stage, wipe with a lint-free cleaning tissue slightly
moistened with water.
6
Position the new adapter plate on the thermal stage. Align the sensor arm with the
corresponding slot on the right side of the thermal stage.
Figure 30
Sensor Arm Location
A
GAIIx adapter plate
B
HiSeq adapter plate
C
Adapter plate sensor arm
7
Tighten the 2 screws to secure the adapter plate. For optimal heat transfer, make sure
that the adapter plate is sitting flat and the screws are tightened evenly.
Summary of Contents for cBot
Page 4: ...iv Material 20004364 Document 15065681 v01 ...
Page 16: ...10 Material 20004364 Document 15065681 v01 ...
Page 22: ...16 Material 20004364 Document 15065681 v01 ...
Page 66: ...60 Material 20004364 Document 15065681 v01 ...
Page 76: ...70 Material 20004364 Document 15065681 v01 ...
Page 88: ...Index 82 Material 20004364 Document 15065681 v01 ...