cBot 2 System Guide
59
Confirm Reagent Delivery (Optional)
You can confirm delivery of individual reagents from the reagent plate provided in any kit
except HiSeq X and HiSeq 3000/4000.
1
Inspect the foil-sealed tops of each tube strip to make sure that each seal was pierced.
2
Release each tube strip from the base of the reagent plate as follows.
a
Grasp the reagent plate firmly, with finger tips under the base.
b
Gently press upward on the center tubes of the tube strip.
3
Inspect each tube to confirm that a similar volume remains in each tube.
NOTE
Small differences are normal.
Figure 27
Example of Successful Reagent Delivery (8-Lane Flow Cell)
Figure 28
Example of Successful Reagent Delivery (2-Lane Flow Cell)
4
If reagent delivery was not successful and the foil seals on the reagent tubes are
pierced, contact Illumina Technical Support.
5
Inspect the 8-tube strip containing library template.
6
If you used additional primers with your run, inspect the 8-tube strip containing
primers.