C O M M A N D R E F E R E N C E
GPIB Command Reference
44
LDC-3926339
C H A P T E R
4
The LASER:CAL:DEFAULT command sets all of the laser calibration constants to default values.
Parameters
None.
Notes
This command should only be used in the case in which the calibration has become corrupted and
circumstances make calibration impossible. Measurements and setpoints on the laser current
source are not guaranteed to meet accuracy specifications after this command is issued. Accuracy
will remain unspecified until calibration is performed on the module.
Examples
Laser:CAL:Default-action: sets all of its laser calibration constants to their default values.
The LASER:CAL:LDI command is used to initiate the calibration procedure for the laser current
setpoint, measurement, and current limit.
Parameters
None.
Notes
Both low and high bandwidth modes, along with current limit, are calibrated during this procedure.
See the Service Reference portion of this manual for complete calibration instructions.
Examples
Laser:CAL:LdI-action: enters calibration mode for laser current.
The LASER:CAL:LDV command is used to initiate the calibration procedure for the laser voltage
measurement.
Parameters
None.
Notes
See the Service Reference portion of this manual for complete calibration instructions.
Examples
Laser:CAL:Ldv-action: enters calibration mode for laser voltage measurement.
The LASER:CAL:MDI command is used to initiate the calibration procedure for the laser photodiode
current setpoint and measurement system.
Parameters
None.
Notes
See the Service Reference portion of this manual for complete calibration instructions.
Examples
Laser:CAL:MDI-action: enters the laser photodiode current calibration mode.
The LASER:CAL:MEAS command is used to enter measured values during the laser calibration
procedure.
Parameters
An <nrf value> which represents the measured value expected for the present Laser calibration
state.
Notes
See the Service Reference portion of this manual for complete calibration instructions.
Examples
Laser:CAL:MEAS 4.95-action: the module uses the value 4.95 for the required measurement value.
LASER:CAL:DEFAULT
LASER:CAL:LDI
LASER:CAL:LDV
LASER:CAL:MDI
LASER:CAL:MEAS