T R O U B L E S H O O T I N G
Hardware Troubleshooting Guide
82
MPS-8033
C H A P T E R
6
!
GPIB Troubleshooting Guide
If you have difficulty operating the 8033 via GPIB, refer to the symptoms listed here.
Output is on but there is no
light at the end of the patch
cord
•
Check the patch cord and the connection with the 8033 and other test
equipment.
Symptoms
Causes and Corrective Actions
No response from 8033
upon a GPIB command
(remote light does not come
on)
•
Check that a GPIB cable is connected between the 8033 and the system
controller (your computer). This cable should be less than 3 meters long.
•
Check that the GPIB address is set properly and the controlling software is
sending commands to the proper address.
•
Make sure no two devices have the same GPIB address.
•
Check that there are less than 15 devices on the bus and there is less than 20
meters of total cable.
•
Check that the GPIB controller card in the host computer is configured
properly.
•
Try isolating the 8033 by removing all other instruments from the bus.
Slow or unexpected
responseto GPIB
commands
•
Make sure no two devices have the same GPIB address.
•
Check that there are less than 15 devices on the bus and there is less than 20
meters of total cable.
•
Make sure the GPIB controller card in the host computer is configured
properly.
•
Try isolating the 8033 by removing all other instruments from the bus.
Device does not respond to
command (remote light is
on)
•
Read the error queue remotely (ERR?). The command syntax or command
structure may be in error.
•
Read the status byte (*STB?) and condition register (COND?) for possible
device problems.
•
Make sure the GPIB controller card in the host computer is configured
properly.
Bus Hangs
•
Place a software timeout around the command in question to prevent the
hang.
•
Check the syntax of the command that is causing the hang−up.
•
Make sure the controlling software is not requesting information from the 8033
without first sending a query command.
•
Make sure the GPIB controller card in the host computer is configured
properly.
•
Check GPIB cables and the GPIB address.
Table 6.1
Hardware Troubleshooting
Symptoms
Causes and Corrective Actions
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for MPS-8033 Series
Page 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 103: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...