background image

   

 

 

Insert the lamps into the lamp holder brackets in the body element. The lamps have pre-
installed fixtures for the gas spring. The fixtures must be positioned at the inside near to 
the camera neck. Picture 20 shows the position of the fixture in the right lamp. 

Note:

  The definition of “right side” and “left side” always refers to the normal operating 

position, i.e. from the front side of the scanner. 

 

 

Picture 20: Position of gas spring fixture at right lamp 

Fix each lamp with two Allen screws M8x12. Use a plastic washer with each screw. 

 

Picture 21: Fixing the lamp 

Connect the two lamp cables to the corresponding connector in the body element. 

 

Page 24 

Setup and Assembly Manual 

Summary of Contents for bookeye 3

Page 1: ...V Ve er rs si io on n R R1 1 Setup and Assembly Manual This device is compliant...

Page 2: ...File BE3 R1 SetupAndAssembly H doc...

Page 3: ...he user operation errors and you can control all functions from the beginning In addition please consider the following points Damages to your unit may have occurred during shipping Please check for d...

Page 4: ...o ensure proper functionality best possible quality and performance of the device This device is a precise optical instrument and should be handled accordingly Section B describes the software setup I...

Page 5: ...ional information concerning measurement of Auto Focus chapter B 2 2 to B 2 2 2 Differentiation between Chassis A and Chassis B H January 2009 New chassis version List of assembling material revised A...

Page 6: ...bling the Components 18 A 2 1 Assembling the Camera Head to the Neck 18 A 2 2 Assembling Lamps and Camera to the Body Element 21 A 2 3 Assembling the Glass Plate 28 A 2 4 Setting the Initial Lamp Posi...

Page 7: ...ocus Measurement with Glass Plate Closed 50 B 2 3 Scan Center 52 B 2 4 White Balance 53 B 2 4 1 Some Basic Information 53 B 2 4 2 White Balance with Glass Plate opened 55 B 2 4 3 White Balance with Gl...

Page 8: ...Target 68 C 3 Recovery Function 69 C 3 1 Important Notes before recovering to Factory Defaults 69 C 3 2 How to Recover to Factory Defaults 69 C 4 Troubleshooting Matrix 71 C 5 Error Codes 73 D Techni...

Page 9: ...ed screws 23 Picture 19 Body element cover opened 23 Picture 20 Position of gas spring fixture at right lamp 24 Picture 21 Fixing the lamp 24 Picture 22 Camera neck assembly 25 Picture 23 Front side 2...

Page 10: ...cture 59 Auto focus measurement screen 51 Picture 60 Scan Center screen 52 Picture 61 Brightness Correction Factor screen 54 Picture 62 R G B controls 54 Picture 63 Glass plate positions for white bal...

Page 11: ...Setup and Assembly Manual Page 11...

Page 12: ...ent back in the original transport box to avoid transport damages The crates and the inserts have been designed and tested to withstand all impacts in accordance with the internationally recognized te...

Page 13: ...ck below the reference folder and two manuals All parts are held in position by foam rubber elements Picture 2 Cardboard separator with lamps and neck Lamps 1 and 2 Reference folder 3 Picture 3 Refere...

Page 14: ...in body and camera head box Two book cradle plates are attached to the scanner s main body Take the cardboard box out of the transport box The cardboard box contains the camera head and the short part...

Page 15: ...amera neck at the scanner s body 8x and to fix the neck to the camera head 2x 4x Lens head Allen screws M4x10 Used to fix the cover of the scanner to the scanner s body 2x Hinges 2x Gas springs for th...

Page 16: ...ensity across the scan area should be kept below 20 If the fluorescent lamps are powered by non electronic ballasts they will produce a flicker twice the frequency of the main power supply 100Hz or 12...

Page 17: ...east three times the weight of the unit Also it should not shake or move to avoid image distortions If the table is too weak it can be attached to a solid wall to stabilize it A 1 2 4 Power outlet Ens...

Page 18: ...a head must be inserted into the neck and must be guided through it One end of the neck and the end of the camera head element are cut at a corresponding angle Insert the cables at this end of the nec...

Page 19: ...nment Picture 8 at both sides To ensure a proper angle alignment between the short neck and the long neck it is necessary to fasten the screws in the given order At first fasten the two Allen set scre...

Page 20: ...elow in the following pictures Picture 11 Camera neck assembled at camera head Picture 12 Placing the cover on the neck Fasten the cover with two Allen screws M8x12 Use plastic washers with the screws...

Page 21: ...out of the transport box must be done by two persons Picture 15 Two persons lift the device out of the transport box Place it on a suitable stable surface The bearing capacity of the table must exceed...

Page 22: ...ly in a place where it cannot be broken Storing it in the transport box is recommended Insert eight Allen set screws M8x8 into the fastening elements where the transport Allen screws were removed Pull...

Page 23: ...pening CAUTION Danger of injury Never put your fingers in the driving element openings Slide the body element cover approximately 300 mm 12 inch to the front side This will make the lamp holder bracke...

Page 24: ...xture in the right lamp Note The definition of right side and left side always refers to the normal operating position i e from the front side of the scanner Picture 20 Position of gas spring fixture...

Page 25: ...old the glass plate bearing Picture 22 Camera neck assembly Note It is recommended to perform the next steps with two persons One person holds and positions the camera neck while the second person han...

Page 26: ...h matching colors Picture 25 Connector board with camera neck cables The connector for the flat ribbon cable has no colored label but is easily identified by its characteristic form Picture 26 Connect...

Page 27: ...ver reaches the edge of the power supply Then release the cable and slide the body cover in its final position Picture 29 Body cover at edge of power supply Insert the screws in the following order 1...

Page 28: ...d after approximately 30 seconds the display shows READY TO SCAN The book cradle plates are controlled by the eight buttons in the middle of the keyboard field Picture 30 Book cradle controller Press...

Page 29: ...position of the axes at the right and left side of the glass plate Picture 32 T slot nut in corresponding position Slide a hinge on each axis Fix each hinge to the lamp with an Allen screw M8x12 Pict...

Page 30: ...p and at the glass plate When the scanner is delivered the gas springs are already mounted to the lamps They only must be mounted to the glass plate To mount the gas springs to the glass plate remove...

Page 31: ...onto the pin of the glass plate Picture 36 Gas spring mounted to glass plate Repeat the procedure with the second gas spring Finally insert the locking springs at both gas springs Picture 37 Locking s...

Page 32: ...nly be done with the help of a second person Open the glass plate completely One person must hold the glass plate in this position while the second person opens the Allen set screw in the fixture slig...

Page 33: ...a reference When a lamp is assembled into the scanner the light beam position relative to the home position switch is unknown To make known all positions of the above described electronic gear the fo...

Page 34: ...ontrol field and simultaneously press the green START button Picture 40 Picture 40 Press the keys simultaneously The lamps light up and move to their start positions The display shows Adjust the light...

Page 35: ...p is changed by pressing the left up down key of the keyboard The position of the right lamp is adjusted accordingly with the right up down key of the keyboard Picture 42 Up down key on right side of...

Page 36: ...eave it in the correct position Picture 43 Light beam on right side of the book cradle Press the down key to move the light beam to the outside Press the up key to move the light beam to the inside Th...

Page 37: ...on to save both lamp positions The sound OK will be played and the display shows Saving new positions The lamp positions will also be saved automatically when the time out period has passed without an...

Page 38: ...the lamps will move at a higher speed than with the up key or down key only In this mode the lamps can be moved to their maximum positions for testing purposes The function of the home position switc...

Page 39: ...refore the address has to be changed To perform this a connection to the scanner must be established First note all network settings in your local PC Then change the settings in your local PC to IP ad...

Page 40: ...level Poweruser using the default password Poweruser Picture 46 Login level screen In the Base Settings section click at Network Configuration Picture 47 Poweruser Main Menu screen Page 40 Setup and...

Page 41: ...cally Picture 48 Network configuration screen If the IP address has been changed the device has to be powered down and up again Change the settings of your local PC back to the previous network settin...

Page 42: ...ensure that adjustments are made following the sequence below 1 Perform the Scan Start function This will synchronize the CCD main drive motor with the lamp motors so that the CCD is centered along th...

Page 43: ...oftware Enter the IP address of the scanner in your browser to open the start menu Picture 49 Scan2Net Start screen In the start screen go to Setup Device Choose the login level Poweruser using the de...

Page 44: ...to Adjustments Picture 51 Poweruser Main Menu screen Start in section Focus Scan Area Adjustments Select the appropriate menu items as described in the following paragraphs Picture 52 Adjustment Supp...

Page 45: ...test target against the neck and center it as precisely as possible Any deviation of the test target in the horizontal direction from the ideal position will be present in all scans afterwards After t...

Page 46: ...text the device is properly calibrated An error is always shown in red text and usually comes with an explanatory remark Picture 54 Scan Start results 1 Click the button New Values in the section Con...

Page 47: ...ition This value measured with the selected glass plate position is the reference for all focus adjustments used later during e g book fold correction folder mode scanning or fixed focus scanning Pict...

Page 48: ...The measurement starts Picture 56 Auto Focus measurement with open glass plate After the Autofocus function has completed the results will be displayed Values displayed in green indicate valid results...

Page 49: ...5 Click the button New Values in the section Controls to repeat the measurement Or 6 Click the button Back to Adj Menu in the section Menu controls to return to the adjustments menu see Picture 52 Set...

Page 50: ...described 9 Move the book cradles to the upper position Close the glass plate Note The position of the book cradles is verified If the book cradles are not in the upper position a Retry button instea...

Page 51: ...ed by some explanatory remarks Note It is normal that the measurement will return different values each time the measurement is repeated The lens motor has a very high resolution and the best focal po...

Page 52: ...d the scanner sees more area at the keyboard side of the scanning bed Since all scanners have some spare pixels that are not used the middle can be adjusted by defining a certain number of pixels at t...

Page 53: ...lmost saturated output signal This assures that the largest density range possible is used After this adjustment is done the uneven light distribution on the CCD caused by the imbalance of the lamps t...

Page 54: ...d way of changing the results is invoked if the individual R G B controls in the scanner application are used They allow changing the color temperature of the original reference scan Picture 62 R G B...

Page 55: ...the Next Step button below the picture corresponding to the selected glass plate position 2 Close the book cradles completely and move them to the lowest position Note The position of the book cradle...

Page 56: ...p button The measurement starts After the Autofocus function has completed the results will be displayed Values displayed in green indicate valid results Any error will be shown in red text followed b...

Page 57: ...is verified If the book cradles are not in the upper position a Retry button and a message are displayed on the screen instead of the Next Step button After moving the book cradles to the upper positi...

Page 58: ...ocus Scan Area Adjustments Press the Laser Check button Note The position of the book cradles is verified If the book cradles are not in the lowest home position a Retry button instead of the Next Ste...

Page 59: ...check result 5 Click the button New Values in the section Controls to repeat the measurement Or 6 Click the button Back to Adj Menu in the section Menu controls to return to the adjustments menu see P...

Page 60: ...er optional functions B 3 1 Time Server Locate the section Base Settings in the Poweruser screen Click the button Adjust Time Date to set various parameters Picture 68 Adjust Time Date screen Select t...

Page 61: ...loaded Picture 69 S2N Sound System screen A sound file can be assigned to various action items The assigned sound will be played every time the condition occurs The default setting can be overwritten...

Page 62: ...update behavior of the scanner from the list Picture 71 Update Scanner firmware screen Browse and select the previously downloaded firmware update file The appropriate firmware can be downloaded from...

Page 63: ...n If the key text does not turn green the key is invalid or does not belong to this specific scanner or option Note Option keys are valid only for one option on a specific scanner denoted by its seria...

Page 64: ...pdates Uploads and go to ICC Profiles Picture 73 ICC Profile screen Select either Printer Profiles or Scanner Profile In the next screen browse for a new ICC profile and select it It will replace the...

Page 65: ...ady slowed down significantly particularly if the file sizes are large Sometimes the network s response times are poor because routers and or switches with slow performance are used or the topology of...

Page 66: ...75 Network Analyzer Parameters screen Check the results in the following screen Single digit millisecond response times are good the example shown is from a well structured gigabit network Picture 76...

Page 67: ...age Quality Server online 24 hours at 7 days a week The Image Quality Server can be reached at http service imageaccess de The server automatically analyzes color tracking white and black points geome...

Page 68: ...ccess de and the system will generate an ICC profile This profile can then be uploaded into the scanner Locate the section Adjustments Support go to Scan Test Targets and click the button Scan IT8 Tes...

Page 69: ...ry Defaults Write down the values for IP address subnet mask and gateway of the device before starting the recovery sequence The recovering to factory defaults described in the following should only b...

Page 70: ...the factory default value 192 168 1 50 8 Change the network parameters to the values which were used before running the recovery sequence Select Setup Device Poweruser Locate the section Base Settings...

Page 71: ...ance Go to the White Balance function and modify the Brightness Correction setting Image is darker on one side than on the other side The electronics gear is out of sync Exercise the Scan Start proced...

Page 72: ...The scanner s optical middle in the horizontal direction is lost or misaligned Exercise the Scan Start procedure A small portion of the lower side of the target is missing in the image The scanner s o...

Page 73: ...ing does not exist The settings does not exist 12 Invalid user docsize The size of the user format is invalid 14 Invalid resolution or color mode Either the resolution or the color mode is invalid 15...

Page 74: ...h cable is defective 30 File format not supported The specified file format is not supported or it is invalid in combination with the color mode 55 E55 WRONG S2N HW CCD PORTS Wrong S2N board detected...

Page 75: ...nnectors to camera 2 69 E69 ADC ERROR CAMERA 3 Camera 3 adc error Test data transfer through analog digital converter failed Check cables connectors to camera 3 75 General keyboard error General keybo...

Page 76: ...LUX Stand by 300 sec 1000 LUX Stand by idle 0 LUX lamps off Lamps High Power White LED UV and IR radiation free Lifetime 50 000 hours on time Book Cradle Maximum load book cradle plate 15 kg D 2 Ambie...

Page 77: ...est mode 150 W Start procedure 260 W Stand by operational lamps off 105 W Stand by operational lamps on 140 W Operating 275 W Pre Scan 170 W Moving the book cradle lamps on Both plates simultaneously...

Page 78: ...1 contains main body lamps camera neck foot pedal test targets assembly material tools and cables 490 x 1055 x 945 mm H x W x D Weight Transport Box 1 89 kg Weight of camera head with short neck elem...

Page 79: ...Notes Setup and Assembly Manual Page 79...

Reviews: