B.2.4.3
White Balance with Glass Plate closed
Start in the
Adjustments & Support
Perform the measurement as follows:
1. Find the section
White Balance Adjustments
and click the button White Balance .
Press the Next Step button below the picture corresponding to the selected glass plate
position. Picture 63 shows the available positions of the glass plate.
2. Close the book cradles completely and move them to the
upper
position.
Note:
The position of the book cradles is verified. If the book cradles are not in the
upper position a Retry button and a message are displayed on the screen
instead of the Next Step button.
After moving the book cradles to the upper position, press the Retry button.
3. Place the
White Reference Target 225x670
at the marked position.
Picture 65: White Balance screen, glass plate closed
4. Press the Next Step button.
After the Autofocus function has completed, the results will be displayed. Values displayed
in green indicate valid results. Any error will be shown in red text, followed by some
explanatory remarks.
5. Click the button Back to Adjustment Menu to return to the adjustments menu.
Setup and Assembly Manual
Page 57