Operation Manual
Page 25
CALIBRATE
The maximum lifting height of the lifting columns is limited by a light
barrier. The light barrier is positioned in the upper part of the bottom unit.
To place the calibration target in the correct position, it must be pressed
to the bottom side of the closed glass plate.
Touch
CALIBRATE
in order to oversteer the light barrier limit. Then lift
the lifting columns until the calibration target is pressed at the glass plate
bottom side.
INIT
Starts the initialization sequence.
Both lifting columns move down to the zero position.
The glass plate closes, i.e. it moves downward completely. Subsequently
it opens to its “INIT” position.
Then both
lifting columns will be lifted to their “INIT” position.
ESC
Closes the parameter menu and returns to the start screen.
TIME SCAN
Defines the time the book cradle and glass plate controller will wait for a
scan command from the scanner. After the time has passed, the
controller moves the book cradles and opens the glass plate.
C.4.1
Touchscreen Parameter Table
The parameter table allows the operator to set value for the initialization and shows status
information.
The first column names the controlled elements.
L1:
Lifting column motor left
L2:
Lifting column motor right
L3:
Glass plate motor left
L4:
Glass plate motor right
UP L1
UP L2
Values from 1 … 100
Defines the speed when moving the lifting columns L1 and L2
upwards.
UP L3:
UP L4:
Defines the motor
’s speed when the glass plate opens.
DOWN L1
DOWN L2
Values from 1 … 100
Defines the speed when moving the lifting columns L1 and L2
downwards.
DOWN L3:
DOWN L4:
Defines the motor
’s speed when the glass plate closes.