Page 28
Operation Manual
D.2
White Balance Adjustment
1. First, p
ress the “INIT” button on the lower Simatic touchscreen to initialize the book
cradle and the glass plate. See picture 8, arrow number. 3.
2. Choose the book mode on the Simatic touchscreen. See Picture 8, arrow 1.
3. Put the calibration plate on the book cradle as shown on the following picture, with
the white calibration target on it.
Picture 10: White calibration step 1
4. Press the F4 button on the Simatic touchscreen to access the parameter menu.
5. Press the calibration button, the calibration mode will then be enabled for 5 minutes.
See Picture 8 arrow 4.
6. Lift left and right half of the book cradle simultaneously to the maximum height by
pressing
the arrow up button “Lift 1-2”.
Picture 11: White calibration step 2