1 INTRODUCTION
Important information: Read the manual carefully before using the device.
Inappropriate utilization may cause damages to the device, to the DUT (device under test)
and/or other components of your setup.
Keep these instructions in a safe place and pass them to any future user.
1.1 Conformity
We, Image Engineering GmbH & Co. KG, hereby declare, that the CAL3-XL corresponds to the
essential requirements of the following EC directive in its current version:
•
Electromagnetic Compatibility - 2014/30/EU
•
RoHS 2 - 2011/65/EU
•
Low Voltage - 2014/35/EU
1.2 Intended use
The CAL3-XL is designed as a calibration light source, based on iQ-LED technology for
cameras with a wide field of view. It includes a micro spectrometer and is controlled with the iQ-
LED control software or via dip switches on the device when not connected to a PC.
•
Only suitable for indoor use.
•
Place your system in a dry and constant tempered environment without any interfering
light.
•
The optimal ambient temperature range is 22 to 26 degrees Celsius. The maximum
ambient temperature range is 18 to 28 degrees Celsius. If the system temperature gets
higher than 38 degrees Celsius, please check your ambient temperature.
•
The optimal system temperature range, displayed in the software user interface, is
between 32 and 38 degrees Celsius. The system has internal temperature
management. If there is any error regarding the internal temperature, you will get a
warning message and the system will automatically shut down to avoid any damage.
1.2.1 Departing from described setup
The following steps must be performed in the correct chronology to allow frictionless
commissioning. Departing from the chronology may lead to an incorrectly working device.
1. Install the iQ-LED software
2. Connect CAL3-XL to the power and then via USB to the PC
3. Switch CAL3-XL on; the system drivers will now be installed
4. Start the software after drivers are installed completely
1.2.2 USB connection
Only appropriate USB connection allows error-free operation of CAL3-XL. Use the delivered
USB cables. If you need to extend the USB connection to longer distances, please check if
powered hubs/repeaters are necessary.