1.5. Technical specifications
1.5.1 General characteristics
Parameters
Value
Working temperature
-25°C / +70°C
Dimensions
132 x 62 x 34mm
Weight
148g 1 Remote probe, 185g 2 remote probes
Casing
IP 68
Radio standards
EN300220-1 et EN300220-2
Network
Sigfox RC1
Max power transmission
14 dBm
1.5.2 Electric characteristics
Supply
Characteristics
Value
Unit
Battery
FANSO EW36mm+51021 connector
Voltage
3.6
V
Current max
150
mA
External
Input range
4.5 to 6.5
V
Current max
110
mA
USB
Tension
5
V
Courrent max
110
mA
The external power supply chosen must comply with the above electrical characteristics and must conform to the standards EN/IEC 62368-1 and
EN 61000-3-2, EN 61000-4-2, EN 61000-4-4, EN 61000-4-5.
Example of reference tested by adeunis and functional with the product by cutting the wires: GS05E-USB - Wall Mount AC Adapters 5W 5V 1A
W/Euro Plug WallMount USB Output from Mean well.
1.5.3 Probes characteristics
Characteristics
Ambient probe - temperature range
[-25°C /+70°C]
Remote probe - sensor temperature range
-55°C..+155°C
Remote probe - wire temperature range
-30°C+105°C
Remote probe - wire length
2m
Precision [0°C/+60°C]
+/- 0,2°C
Precision [-35°C/0°C]
+/- 0,5°C
Precision [-55°C/-35°C]
-0,6 /+0,8°C
Precision [+60°C/105°C]
+/- 2°C
CAUTION: Allow a few minutes for the probe to come back to room temperature before any manual operations to avoid mechanical damages.
In addition, please respect the minimum and maximum temperatures for the sensor and the wire (cf. Table here-above).