Echo
®
Wireless Vibration Sensor Manual
Version 1.2, 2011-12-07
16
Mounting
The Echo
®
Wireless Vibration Sensor should be stud mounted on a clean, flat surface. It
is not recommended to mount the sensor with a magnet due to the size and mass of the
sensor. The correct mounting torque for the sensor is 3 to 5 ft-lb (4 to 7 N-m). It is also a
good idea to put a small amount of silicon grease on the bottom of the sensor before
mounting.
WARNING
– Do NOT tighten the sensor by twisting on the housing. Damage to the
sensor WILL occur. Tighten to the correct torque using a wrench on the hex base.
Outline Drawing – See enclosed drawing