5
1.1.10 Technical
Data
Power supply:
7V..15V-DC, 600 mA/h battery capacity
Supply current:
8mA standby
200mA while 10..15sec. measuring time
Resolution: 0,1%
Repeating accuracy:
±
0,3%
Measurement range
0-100%
Temperature range:
-15°C...50°C, expanded temperature ranges on request!
Temperature caused value drift:
max.
±
0,5%
Standard interface:
analogue output 0..1V or 0(4)..20mA, RS232/V24,
IMP232
MICRONET
Calibration:
calibrated for mineral soils,
but individually adaptable per software
Case:
weatherproof, robust aluminium die-cast (IP65)
The measuring accuracy and -range as well as the tolerable range of bulk electrical conductivity
depends on the corresponding probe.
1
2
5
6
7
3
4
Pin 1: Reserved
Pin 2: Reserved
Pin 3: Analog-Out 0..1V
Pin 4: Analog-Out-0V
Pin 5: SelfCal.
Pin 6: R/T (IMP-Bus)
Pin 7: COM (IMP-Bus)
Coax cable
RS2
/
32 V24
Probe connector
Connector for
- analogue cable
- calibration plug
- battery charger
AC-Connector
AC-banana plugs
Battery
charger
external
battery
box
Connector for
- external power supply
- RS232/V24
front view
Pin 1: +U (9V..15V-DC)
max. 250mA
Pin 2: RxD
Pin 3: 0V
Pin 4: TxD
f ront vi ew
1
2
3
4
Probe
P2
Probe
P2D
Probe
P2Z
Connectable probes for TRIME-FM2
Connectable probes for TRIME-FM3
Probe
P2G
Tube Probe
T3
Probe
P3
Probe
P3Z
Start
Probe
P3S
Po
we
r
su
pp
ly
Summary of Contents for TRIME-FM