Place the front case assembly over
the vent module and tighten the (2)
4-40 X ¼ screws unto the dovetail
mounting bracket. Reconnect the
ribbon cable unto the PIM PCB.
Make sure the two locking “ears”
lock into position.
18
Attach the back case to the vent
module and align cover with handle,
air intake housing and dovetail
mounting studs. Insert and tighten
the included (2) 10-32 Keps nuts and
the (4) 6-32 X 2” screws.
19
Tighten the (4) 8-32 X 3 screws on
the outer air intake.
20
Perform HiPot Testing then perform
Calibration and Functional Test using