906-0731-04 Rev. B Sept. 2012
Page 29 of 68
Power Knob Kit 712-0731-06 / 712-EGL2-06
Contents:
Instructions:
STEP NO.
DIRECTIONS
1
Loosen but do not remove the
(4) 8-32 X 3 screws on the outer air intake.
2
Loosen and remove the (2) 10-32 Keps nuts
and the (4) 6-32 X 2 screws. Remove the
back case by lifting from the ventilator.
3
Remove the (2) 4-40 X ¼ screws on the
Dovetail Mounting Bracket and disconnect
the ribbon cable on the PIM PCB by
simultaneously applying pressure on the
two locking “ears”.
4
Flip the ventilator over and remove the
front case assembly by lifting it straight up
away from the ventilator module.
5
Remove the bezel by loosening and
removing the (7) 4-40 X ¼ screws that hold
the bezel to the front case.
ITEM
QTY
DESCRIPTION
1
Power Knob
1
Screw, Socket, Cup Point, Set, 6-32 X 1/4 (installed into knob)