906-0731-04 Rev. B Sept. 2012
Page 35 of 68
5
Place the replacement front case
assembly over the vent module and
tighten the (2) 4-40 X ¼ screws unto
the dovetail mounting bracket.
Reconnect the ribbon cable unto the
PIM PCB. Make sure the two locking
“ears” lock into position.
6
Attach the back case to the vent
module and align cover with handle,
air intake housing and dovetail
mounting studs. Insert and tighten the
included (2) 10-32 Keps nuts and the
(4) 6-32 X 2” screws.
7
Tighten the (4) 8-32 X 3 screws on the
outer air intake.
8
Select the correct label for your
device, peel the backing off and affix
to the front bezel.
9
Update serial numbers using
the RCS:
a.
Use the “C:\program
files\impact\EMV\bin\
EmvDLGui.exe” to change
the serial numbers.
b.
Use the RCS FD app to re-
flash the firmware.
10
Perform HiPot Testing then perform
Calibration and Functional Test using
the RCS.