906-0731-04 Rev. B Sept. 2012
Page 59 of 68
Oxygen Inlet Fitting Kit 712-0731-18
Contents:
Instructions:
STEP NO.
DIRECTIONS
1
Remove the damaged O2 Inlet fitting by
unscrewing the dust cap then unscrewing the
(3) 8-32 X 7/16 screws.
2
Install the included O-Ring into groove on
underside of inlet fitting.
3
Place Oxygen Inlet Fitting over the ventilators
connector panel then insert and tighten the 3
supplied 8-32 X 7/16 screws.
4
Perform HiPot Testing then perform
Functional Test using the RCS.
ITEM
QTY
DESCRIPTION
1
Oxygen Inlet Fitting with Filter, Cap and Chain
3
Screw, Phillips, Flat Head, 8-32 X 7/16, ZP
1