906-0731-04 Rev. B Sept. 2012
Page 64 of 68
2020
O2 Supply Pressure Low (O2 Tank
Pressure Low)
O2 supply pressure: trigger < 35
psig, clear > 40 psig
1030
Fresh Gas Intake Failure
Fresh Gas / Emergency Air Inlet
blocked
712-0731-02
2030
Fresh Gas Intake Fault (Compressor
Intake Blocked - Backup Available)
Fresh Gas / Emergency Air Inlet
blocked
3030
Fresh Gas Intake Fault (Compressor
Intake Blocked - Backup Selected)
Fresh Gas / Emergency Air Inlet
blocked
3031
Fresh Gas Intake Fault (Compressor
Intake Restricted)
Fresh Gas / Emergency Air Inlet
blocked
712-0731-02
3032
Fresh Gas Intake Fault (Intake Pressure
Signal Chain Failure)
communication between Fresh
Gas / Emergency Air Inlet
pressure sensor and SPM lost
712-0731-02
1041
O2 Supply Pressure High Failure (O2
Tank Pressure Excessive - No Backup)
O2 supply pressure > 80 psig
Lower Oxygen
input pressure
3041
O2 Supply Pressure High (O2 Tank
Pressure High)
O2 supply pressure: trigger > 75
psig, clear < 68 psig
1051
Run-Time Calibration Failure
failure of calibration system
712-0731-02
1052
Airway Pressure Sensing Failure
communication between airway
pressure sensor and SPM is lost
712-0731-02
2053
Suspicious Triggers (False Trigger or
Bad Baseline)
airway pressure sensor fails to
calibrate during expiratory
phase of breath
712-0731-02
1060
Exhalation System Failure (Exhalation
Valve Failure)
exhalation control valve fails to
operate
712-0731-02
1061
Exhalation System Failure (Excessive
Airway Pressure)
triggers (airway pressure): > 40,
75 cmH2O
712-0731-02
2062
Exhalation System Fault (Gas Trapped)
airway pressure at end of
expiration > 5 cmH2O above
PEEP
712-0731-02
2070
Airway Pressure High
Paw > high limit
712-0731-02
2071
Low Airway Pressure
Paw < low limit
712-0731-02
2072
High Tidal Volume
Vt > high limit
Check
internal/exter
nal tubing
connections
2073
Low Tidal Volume
Vt < low limit
2074
High Breath Rate
BPM > high limit
2075
Low Breath Rate / Apnea
BPM < low limit
2076
Apnea
BPM < low limit
2090
PEEP Leak
Paw < [PEEP - 2 cmH2O] during
expiration
3091
Incomplete Exhalation
exhaled flow from patient
continues throughout expiratory
period
3092
Patient Inspiratory Demand Not Met
end inspiratory Paw < -1.0
cmH2O
2095
Insufficient Flow
pressure target not reached
during inspiratory period during
pressure-targeted ventilation