906-0731-04 Rev. B Sept. 2012
Page 66 of 68
2314
Pulse Ox Sensor Off Patient
3315
Pulse Ox Light Contamination
3316
Invalid Pulse Ox Sensor
3317
Low SpO2 Perfusion
3318
Low SpO2 Perfusion
2401
SpO2 Low
2410
Heart Rate High
2411
Heart Rate Low
1420
Complete Power Failure
2421
Input Protection Circuit Failed
failure of input protection circuit
702-0731-02
3421
External Power Fail/Disconnect
external power < 5 VDC
3422
Missing Battery
communication between
battery and CPU has failed
703-0731-01;
702-0731-02
2423
Power Circuit Hardware Fault
internal power circuit has failed,
external power connected but
cannot be used
702-0731-02
3423
Battery Charge Circuit Failed
1430
Empty Battery
703-0731-01
2430
Low Battery (Low Battery - No Backup)
< 5 min of battery operation
remaining
3430
Low Battery (Low Battery - Warning)
< 30 min of battery operation
remaining
3431
Low Battery (Low Battery - With
Backup)
< 30 min of battery operation
remaining
703-0731-01
3441
External Power Failed (External Power
High)
external power: trigger > 33
VDC, clear < 30 VDC
702-0731-02
3442
External Power Failed (External Power
Low)
5V < external power < 11.5V
702-0731-02
3444
External Power Failed
external power voltage polarity
reversed
Use Impact
Power supply
Only
2450
Battery Fault - No External Power
Connected (Battery Nearly Too Hot for
Discharge)
battery temperature > 70C
703-0731-01
3450
Battery Fault - With External Power
Connected (Battery Nearly Too Hot for
Discharge)
battery temperature > 70C
3451
Battery Fault - With External Power
Connected (Battery Too Hot for
Discharge)
battery temperature > 75C
3452
Battery Fault (Battery Too Hot for
Charging)
battery temperature > 45C
3453
Battery Fault - With External Power
Connected (Battery Too Cold for
Charging)
battery temperature < 0C
2455
Battery Fault - No External Power
Connected (Communication Failure)
EMV not able to communicate
with internal battery
703-0731-01;