906-0731-04 Rev. B Sept. 2012
Page 67 of 68
3455
Battery Fault - With External Power
Connected (Battery Communication
Failure)
EMV not able to communicate
with internal battery
3470
Internal Communication (Comm)
Failure Fault - PIM Comm
EMV not able to communicate
with PIM
Check Cable
from PIM
board to CPU;
702-0731-02
1471
Internal Communication (Comm)
Failure
CPU unable to communicate
with UIM and interface controls
Check Flex
cable to LCD;
712-0731-14
1472
Internal Communication (Comm)
Failure
CPU unable to communicate
with SPM
Check Cable
from PIM
board to CPU;
702-0731-02
1473
Internal Comm Failure
no valid data sent from SPM
Check Cable
from PIM
board to CPU;
702-0731-02
1474
Internal Comm Failure
CRC check between EMV and
SPM fails
Check Cable
from PIM
board to CPU;
702-0731-02
1475
LCD Control Failure
CPU unable to control LCD
contrast
712-0731-08
1480
SPM Compatibility Failure
EMV and SPM software revs
incompatible
Recalibrate
using RCS
3480
SPM Compatibility Fault
EMV software detects that it has
not been calibrated with SPM
inside unit
1485
Power-On Self-Check Failure
SPM software fails and is shut
down
712-0731-02