906-0731-04 Rev. B Sept. 2012
Page 45 of 68
Back Case Kit 712-0731-12 / 712-EGL2-12
Contents:
Instructions:
STEP NO.
DIRECTIONS
1
Loosen but do not remove the
(4) 8-32 X 3 screws on the outer air intake.
2
Loosen and remove the (2) 10-32 Keps nuts
and the (4) 6-32 X 2 screws. Remove the
case by lifting from the ventilator.
3
Place the replacement cover over ventilator
and align cover with handle, air intake
housing and dovetail mounting studs. Insert
and tighten the included (2) 10-32 Keps nuts
and the (4) 6-32 X 2” screws.
4
Tighten the (4) 8-32 X 3 screws on the outer
air intake.
5
Perform HiPot Testing then perform
Functional Test using the RCS.
ITEM
QTY
DESCRIPTION
1
Back Case Assembly
4
Screw, Phillips, Pan Head, SS, Black Oxide, 6-32 X 2
2