165910 Dew Point Meter 8072A Manual
11
June 2014
The 8072A is factory set to
display dew point in
q
F. The
display can be changed to
q
C by jumper configuration.
With the 8072A panel
removed from the carrying
case, locate the
configurable parts near the
upper left corner of the PC
board. Move J6 and J7
jumpers to the C position to
display dew point in
q
C.
To change the dew point at which the monitor indicates an
alarm (high dew point indication), use the DISPLAY MODE
SWITCH labeled S1 and the ALARM ADJUST potentiometer
labeled SET1.
WARNING: The alarm set point adjustment is performed with
the 8072A powered on. Avoid touching live electrical parts
during this procedure as hazardous voltages are present!