Page 1: ...LUND DCT6LI FIELD MANUAL AND TROUBLESHOOTING GUIDE ...

Page 2: ...ooting Guide Contents Overview of Terms 2 Power System 5 Power System Access 5 PowerVar Inverter 6 Power Sonic LiFe Battery 8 Components 9 Lift Mechanism 9 Power Cord 10 Troubleshooting 11 Basic Troubleshooting 11 Table of Condition codes 14 ...

Page 3: ...System Controller Remote User Interface RUI Slide out Keyboard Tray Optional Drawer Combi Cam Drawer Lock 75 100mm VESA Display Mount Accessory Mounting Holes Combi Cam CPU Compartment Lock AC Cord Holder Optional Wire Basket ...

Page 4: ... outlet from Power module Remove 4 8 32 Screws for wire access USB Comm cable for power module Lift control switch Lift control switch wire to lift controller AC Coil Cord from power module Note Install from the top down ...

Page 5: ...Universal mounting rail Optional wire basket Adjustable keyboard tray for flat or ergonomic inverted positions ...

Page 6: ...er System Access To access the battery power cord power module and wiring remove the 6 Philips head screws surrounding the base The power cord can be replaced after removing the rear screw and lifting the access panel ...

Page 7: ...PowerVar Inverter The inverter is secured to the base of the cart by 4 Philips head screws at the base ...

Page 8: ...The front panel has the termination points for the power status panel the battery communications cable and the power connections to and from the battery and cart ...

Page 9: ...ecured by the foam that it is set in The wiring to the battery terminals should be disconnected before the battery can be removed The red fuse holders contain a 40A fuse between the battery and power module and a 10A fuse between the battery and lift control relays ...

Page 10: ...The lift control board is located in the box underneath the worksurface on the back of the lift column Lift motor connector P1 30A Motor control relays Lift Control switch connector RJ45 P2 Red Black 12V power from LiFe battery ...

Page 11: ...Power Cord The power cord can be easily replaced after removing the access panel screw at the rear of the cart base Compress the cable retention insert and pull it out ...

Page 12: ...id Battery and Invalid Input AC Check wall socket and test for proper line voltage MPM input power cord is not plugged in Plug in input power cord Output fuse is open Reduce load replace fuse and test Backup time is less than expected Battery is not fully charged or battery is worn out Recharge battery for at least 24 hours and retest backup time MPM seem normal but the load will not turn on power...

Page 13: ...also cause the Clinic View application to open automatically to notify the cart user of the time remaining and charge level Service Related Conditions Service related alarms such as overload over temperature replace battery etc will cause one of the RUI Service LEDs to flash The specific alarm that is raising the Yellow or Red status will be logged and displayed locally in the Tech View applicatio...

Page 14: ...be defined at a pack level by a particular manufacturer Typically it will be in the range of 20 C to 75 C Remaining Capacity RemainingCapacity 0x0f The predicted remaining battery capacity Cycle Count CycleCount 0x17 The number of cycles the battery has experienced A cycle is defined as An amount of discharge approximately equal to the value of DesignCapacity Battery Status BatteryStatus 0x16 A li...

Page 15: ...city is less than x of original design capacity measure of state of health Defaujlt is 50 configure the threshold via MPM View 24 063 Service Check Replace Battery Date Threshold Raised when the comparison of MPM View system date to Battery Replace Date indicates that the battery Age is older than the Battery Age Threshold Default is 18 Months configure via MPM View 24 064 Service Check Smart Batt...

Page 16: ... Charger Raised if MPM charger delivers more than 110 of max charge current for more than 60 seconds 33 038 Service Required Charger Raised when MPM is attached to a Smart Battery and battery indicates an Over Charged alarm 34 053 Battery Charge Threshold Low Critical Final warning to recharge battery shutdown imminent Hard coded within MPM to be raised when charge level is less than 5 34 055 Serv...