VuSitu Calibration & Settings
About Calibration and Settings
You can perform sensor calibrations, view a calibration report, or restore factory calibration defaults when the instrument is
connected to VuSitu.
1. Tap the VuSitu menu icon, and select Connected Instrument from the list.
2. Tap Calibrations.
3. Calibrations and settings for all connected sensors will appear. Ammonium, chloride and nitrate parameters are not
shown in the image below but are displayed when the corresponding sensors are installed.
4. Tap the calibration you want to perform.
You can also access the full Calibration Report from this menu.
Quick-Cal Multiple Sensor Calibration
Quick-Cal allows you to perform a one-point calibration on up to three sensors with one setup and stabilization process.
1. From the main menu, select
Connected Instrument
.
2. Select
Calibrations
.
3. From the Calibration Menu, select
Quick-Cal (multi-sensor)
.