18
Connecting a Pulse Instrument to VuLink
VuLink can upload data from a pulse instrument to HydroVu or an FTP server.
Refer to the following table when connecting a pulse instrument to VuLink via a Rugged Cable.
Required Components
Wiring with the Load-Bearing Universal Adapter (LBUA)
• Rugged Twist-Lock Cable with one stripped-and-tinned end
• Load-Bearing Universal Adapter (LBUA)
• Cable (from LBUA to pulse instrument)
• Pulse instrument
• VuLink
Rugged Cable Wire Legend
Wire Color
Signal
Brown
Pulse Output
Black
Ground
Red
Power (optional)
Blue
Unused
Green
Unused
White
Unused
Attach the Twist-Lock
end of a Rugged Cable to
VuLink.
Connect the brown and
black wires at the other
end of the Rugged Cable
to the Load-Bearing
Universal Adapter.
Snip the four unused
wires.
Run wires from the other
end of the LBUA to the
pulse instrument.
To VuLink
To pulse
instrument
1-970-498-1500
www.in-situ.com