4
5
B. Operating Environment
Please make sure that the environment is adequate for the InBody370S installation. This equipment is designed
for indoor use. If installing outdoors, the following requirements must be fulfilled.
C. Installation Instructions
1. Open the packing box of the InBody370S and remove the packing pads. Then take the InBody370S out of the box.
2. After completely raising the upper part of the InBody370S, remove the polyethylene foam cover.
Temperature range
10 ~ 40°C
(50 ~ 104°F)
Relative humidity
30 ~ 75% RH
Atmospheric pressure
70 ~ 106kPa
* Please refer to the following illustrations to properly transport the equipment.
Polyethylene foam cover
Packing pads
Joint frame
•
If you have any problems installing your InBody370S, please contact InBody for assistance.
•
Do not transport the equipment by holding the screen portion or the joints of the hand electrodes.
•
Keep the packing materials provided for repacking the equipment in the future. Other wastes should be
disposed of according to relevant laws and regulations.
Caution
•
Using the InBody370S on carpet may cause static electricity, which could damage the equipment. If installing
the InBody370S on carpet is unavoidable, please use an antistatic mat.
•
Install the InBody370S on a leveled, non-vibrating surface. Installing the equipment on an uneven surface
may cause the examinee to fall down. Test results may also be inaccurate.
•
Never clean the hand and foot electrodes with liquid spray or detergent directly. The equipment may corrode
and/or malfunction if the liquid or detergent leaks inside. Use the InBody Tissue when cleaning the InBody370S.
Caution
370S_manual_Eng_C_181224.indd 5
2018. 12. 24. 오후 7:29