  Connect the sensor cable to the gauge. 

 

 

  Secure the cable connector with the lock screws. 

  Connect the sensor cable to the controller. 

 
 
 
 

 

Operation 

When the voltage is supplied, the measuring signal is avail-
able between pins 2 (+) and 12 (–) (Relationship Measuring 
Signal – Pressure 

 "Technical Data" and 

 [1]). 

BPG400-SD, -SP and -SR can also be operated via the 
corresponding fieldbus interface (DeviceNet, Profibus or 
RS485 

 

 [1] and [2] for further details and functions). 

Allow for a stabilizing time of 

10 minutes. Once the gauge 

has been switched on, permanently leave it on irrespective of 
the pressure. 
 
 

 

Gas Type Dependence 

(BPG400)

 

The measurement value is gas dependent. The displayed 
reading applies to dry air, O

2

, CO, and N

2

. For other gases, it 

has to be converted (

 "Technical Data" and 

 [1]). 

 
 

 

Adjusting the Gauge 

 

  The adjustment of BPG400-SD, -SP and -SR 

(

 

 [1] and [2]) is slightly different from the 

procedure for BPG400, which is described below. 

 

The gauge is factory calibrated. If used under different 
climatic conditions, at extreme temperatures, through aging 
or contamination and after exchanging the sensor, the char-
acteristic curve can be offset and readjustment can become 
necessary. Only the Pirani element can be adjusted and only 
at atmosphere. 

Readjustment becomes necessary if 

  at atmosphere the output voltage is <10 V or the display 

reading is <atmosphere 

  when venting the vacuum system, the output voltage 

reaches 10 V before the measured pressure has reached 
atmosphere (Gauges with display will show the error "5" at 
atmosphere (Pirani sensor warning)). 

 

  Activate the gauge. 

  Operate the gauge for 

10 minutes at atmos-

pheric pressure. If the gauge was operated 
within the BA range, a cooling-down time of 

30 minutes is to be expected (gauge tem-

perature = environmental temperature). 

 
 
 
 
 

  Adjusting the gauge 

 

BPG400 without display 

353-500   353-502 

BPG400 with display 

353-501   353-503 

Insert a pin (

ø1.3×50mm) through the opening marked 

<FULL SCALE> and push the button inside for at least 

5 seconds. 

 

 

 

 

10 s 

Automatic adjustment 

 

 

 

Adjustment completed 

 

 

 
 

 

Display 

(BPG400 with part numbers 353-501 and 353-503) 

 

Pressure reading

Pressure unit

 

 

Function display 

(none) Pirani operation

Emission 25 

μ

A

Emission 5 mA
Degas
1000 mbar adjustment (Pirani)

 

 

Error display

 

no error

(green background illumination)

Pirani sensor warning

(red background illumination)

Pirani sensor error

(red background illumination)

 

 

 

 

BA sensor error

(red background illumination)

 

 
 

 

Internal data connection failure 
(red background illumination) 
 
 
 
 

 

Deinstallation 

 

 

DANGER

 

DANGER: contaminated parts 
Contaminated parts can be detrimental to health 
and environment. 
Before beginning to work, find out whether any 
parts are contaminated. Adhere to the relevant 
regulations and take the necessary precautions 
when handling contaminated parts. 

 

 

 

 

Caution

 

Caution: vacuum component 
Dirt and damages impair the function of the vac-
uum component. 
When handling vacuum components, take ap-
propriate measures to ensure cleanliness and 
prevent damages. 

 

 

 

 

Caution

 

Caution: dirt sensitive area 
Touching the product or parts thereof with bare 
hands increases the desorption rate. 
Always wear clean, lint-free gloves and use 
clean tools when working in this area. 

 

 

 

  Vent the vacuum system. 

 

  Put the gauge out off operation. 

 

  Unfasten the lock screws and unplug the sensor cable. 

(If you are using BPG400-SD, -SP or -SR, unfasten 
and unplug the interface cable too (

 

 [1] and [2]). 

 

 

 

  Remove the gauge from the vacuum system. 

 

Protective lid

 

 