Easy Rate Sensor
PN 074-677-P1A
3
Using Teflon
™
tweezers, gently pry the crystal holder spring out of the sensor body.
(See
Figure 10 Crystal holder spring
NOTE:
Move the spring circularly until the gap between the two ends is found and
pry it out from there.
4
Loosen the two 0-80 screws and remove the 0-80 retainer plate. (See
Figure 11 Loosen screws and retainer plate
5
Pull the ceramic retainer assembly from the sensor body and inspect the cavity in the
sensor and components for debris or any possible contamination. If observed, clean
the sensor body and components.
6
Solder the wire onto the microdot connector. Prior to soldering, the wire end may need to
be crimped to fit the microdot slot.
For standard sensors, the wire should be soldered in parallel to the connector as
shown in
Figure 12 Standard sensor wire positioning