26
tina84e1-a (2019-07)
Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). or …
max. ø1.1 mm
Press the
button briefly
… apply supply voltage to Remote Zero at pin 10
(pulse
→
The zero adjustment runs automatically. The <STATUS>
indicator flashes until the adjustment (duration
≈
8 s) is
completed.
After zero adjustment, the gauge automatically returns
to the measurement mode.
The <STATUS> LED blinks green if
•
the signal output is negative (< -20 mV) when the final pres-
sure has been attained
•
the zero adjustment has failed.
4.2.2
<
ZERO
>
Adjustment with Ramp Function
The ramp function allows to adjust the pressure value at a
known reference pressure within the measurement range of the
gauge.