tina84e1-a (2019-07)
35
4.5
Diagnostic Port (USB Interface)
The diagnostic port <DIAG> permits to output the pressure
reading and all status information and to enter all settings at the
same time (
→
[4]). A standard USB cable type A/mini B can
be used.
Required software: T-Gauge. Please contact your local INFICON
service center.
Mini USB type B
4.6
EtherCAT Operation
Caution
Data transmission errors
Any attempt to simultaneously operate the gauge
via the EtherCAT interface and the diagnostic port
causes data transmission errors.
Therefore the gauge must not be operated simul-
taneously via the EtherCAT interface and the diag-
nostic port.