Analog Measurement Signal vs. Pressure
1
2
3
4
5
6
7
8
9
10
0
0.9×F.S.
0.5×F.S.
0.6×F.S.
1.0×F.S.
0.7×F.S.
0.8×F.S.
0.1×F.S.
0.2×F.S.
0.3×F.S.
0.4×F.S.
1.1×F.S.
Pressure p
Measurement U
out
[V]
0.0×F.S.
p = (U
out
/ 10 V) × p (F.S.)
Conversion Torr
↔
Pascal
Torr
mbar
c
1.00
1013.25 / 760 =
1.3332…
101325 / 760 =
133.3224…
Example: Gauge with 10 Torr F.S.
Measurement signal U
out
= 6 V
p = (6 V / 10 V) × 10 Torr
= 0.6 × 10 Torr =
6 Torr