18
Warning
In rare cases, the sensor electrode may fracture.
Don't ignore it. If the sensor electrode breaks and
there is no visible sensor electrode on the skin, do not
attempt to remove it. Seek professional medical help
in the condition of infection symptoms or inflammation
- redness, swelling or pain. If the sensor electrode
ruptures, please report to our technical support
department.
No safety hazard was found in the patient's
residual rupture sensor electrode during in vitro NMR
test. There was no obvious migration or heating of the
sensor electrode, and the imaging was limited to the
area around the sensor electrode.
6.1 Preparation
Test the blood glucose according to the instructions of the
blood glucose meter to ensure the accuracy of the blood glucose
value during calibration.
Wipe the bottom of the transmitter with a wet cloth or
alcohol wool. Place the transmitter on a clean, dry cloth and air
dry for 2-3 minutes.
6.2 Transmitter installation
After the sensor electrode is inserted, the transmitter must
be fixed to the sensor base. The steps are as follows:
1.Wipe the bottom of the transmitter with a wet cloth or
alcohol wool and air dry. Do not touch the bottom of the
transmitter, as the metal contacts is on the bottom. Do not
scratch the bottom of the transmitter, as the scratch may
damage the waterproof.
2.Place the transmitter in the sensor base.
3.Fix the transmitter
a. The thumb and middle finger of one hand are pressed
on the edges of the adhesive tape and the base, and the index
finger is placed on the transmitter to ensure that it will not
move.
b. Press the transmitter with the thumb of the other hand.
c. When the transmitter is fixed, make sure you hear the
tick. If it is not completely buckled, it may cause the transmitter
to disengage.
Figure 20 Fix the transmitter