Contraindications
Taking acetaminophen while wearing the sensor,
blood glucose readings may be higher.
7.1 Overview
The glucose readings need to be calibrated to display the
readings and trend curves, and to maintain the accuracy
throughout the monitoring process. The time points for
calibration are as follows:
1.First calibration: After the sensor is activated and the
sensor polarization is complete, App prompts the reference
blood glucose for the first calibration. The App will not display
the blood glucose readings and trend curves until the first
reference blood glucose is input for calibration.
2.12- hour calibration: the reference blood glucose should
be re-entered every 12 hours for calibration after the first
reference blood glucose input. The purpose of calibration is to
keep the blood glucose readings more accurate.
3.Need more information or other reasons.
Caution
Blood glucose must be entered manually during
calibration. It must be calibrated by accurate blood
glucose values in order to obtain accurate blood
glucose readings.
7.2 How to calibrate
Accurate blood glucose values need to be entered for each
calibration. Blood glucose for calibration must be between 2.2
and 22.2 mmol/L and must be within 3 minutes of reading.
1.If the blood glucose measured by the glucose meter is
outside the range of 2.2-22.2 mmol/L, it cannot be used for
calibration.
2.Before calibration, make sure the transmitter and mobile
App are connected.
3.Use a blood glucose meter that routinely tests blood
glucose for calibration. Do not replace the glucometer during
the monitoring process. The accuracy of different brands of
blood glucose meters and test strips will vary.
4.The accuracy of blood glucose readings used for
calibration may affect the accuracy of blood glucose readings.
20