26
At this point, if the same type of prompt occurs, it will not be
prompted again within the repeated prompt time. If the current
repeated prompt is not manually canceled or the reminder
condition still exists, the App will issue a prompt according to
the repeated prompt interval. If the prompt is not manually
canceled after the first prompt is issued, it will be prompted
every 3 minutes.
In the alarm notification interface, click " repeat interval"
to set the repeat time.
9.3 No disturbing
In certain fixed time periods, such as sleep, notification is
not wanted which will affect life or rest, so you can turn on the
function of no disturbing. In the notification interface, click
[Do Not disturb duration] to set the start time and end time ,
and click the [Do Not disturb] icon to turn on the function.
When the function is enable, the App will not make any sound
or vibration even when prompted.
9.4 Disconnect notification
When the transmitter and mobile phone are far away or
obstructed, the App and the transmitter will be disconnected.
The connection mark of the main interface is used to indicate
the connection status of the App and transmitter. Manually
clicking the connection icon can switch the connection state
between App and transmitter.
You can enable or disable the disconnection prompt function
on the notification interface. In the notification interface, click
[Disconnect notification] to open the status and enable the
Disconnect notification function.
9.5 Hyperglycemia notification
The default value of hyperglycemia is 11.1 mmol/L.
Click the [high Alert] icon to the open state and set the
prompt value, which can be manually adjusted between 6.7-
22.2mmol/L. When the blood glucose reading is higher than
the prompt value, the APP will trigger the prompt and remind
it with sound or vibration.
9.6 Hypoglycemia notification
The default value of hypoglycemia is 4.4 mmol/L.
Click the [Low Alert] icon to the open state and set the
prompt value, which can be manually adjusted between 3.3-
5.5mmol/L. When the blood glucose reading is lower than the
prompt value, the APP will trigger the prompt and remind it
with sound or vibration.
9.7 Rising rate notification
The default value of rising rate is 0.2 mmol/L/min.
Click the [Rise rate Alert] icon to the open state and set
the prompt value. When the change rate of blood glucose is
higher than the prompt value, the APP will trigger the prompt
and remind it with sound or vibration.