V1.0.1 - 9 -
Yantai IRay Technology Co., Ltd.
11th Guiyang Street, YEDA
Yantai 264006, P. R. China
When selecting the reference blackbody in the operation interface, it is recommended
to select the 1/2 center area of the reference blackbody surface;
D) In the process of temperature measurement, it is better to make sure that the
measured target (person) is in the same cross section with the reference blackbody.
E) It is recommended to put the blackbody at the position of 1.5m away from camera.
2) Reference black body layout
As shown in Figure 9, adjust the direction and position of the infrared human body
thermal imaging camera and the reference black body so that the reference black body
appears in the 2/3 central area of the camera market and shall not be shielded by
pedestrians.
Figure 9
Infrared Imaging on PC
⚫
Environmental requirements
The infrared temperature measuring camera is arranged in an environment where the
temperature is relatively constant and there is no direct sunlight, to avoid temperature
measurement errors caused by the temperature changing environment affecting the
human skin surface temperature. According to the above installation requirements, the
following installation methods can be used to install equipment in halls and aisles.
3.5 Site Installation and Layout Considerations
1
)
Avoid placing it at the entrance with airflow disturbance
Avoid the temperature rise and fall caused by the air surface disturbance of the skin of
the person being tested, which will affect the false alarm or miss alarm of the infrared
temperature measuring camera.
参考黑体
出现区域
人
Reference blackbody
appearance area
People