INFLUX BRASIL – Av Paraná, 3059 – Cajuru do Sul — Sorocaba
– São Paulo - CEP:18105-002 Tel.: (15) 3359 9739 (15) 99621
1260 - e-mail: vendas@influxbrasil.com
4.4 Parameter setting
4.4.1 Calibration steps
1
)
Liquid crystal display the level value after instrument plug in,
shown as the right figure, the actual level is 3.320m, calibration method
described as following
2
)
Using the “inductive-pen” to click Mode key to enter the parameter
setting menu, first entry is
the P01(level)calibration menu.
3
)
Click the OK button, enter the level calibration input interface, input
actual value of 3.320m, as the right figure, click Mode button return to
level menu
4
)
Then click Mode button again to display as shown, click Mode button
to return to parameter setting menu, click OK button to exit.
5
)
The instrument displays the actual level, from then on, the meter
readings will accurately reflect the actual level value.
4.4.2 Parameter 4~20mA setting
1
)
In a normal operation, click Mode button to enter the parameter
setting menu, then click▼to choose P02(4
~
20mA) setting menu,
shown as right figure.
2
)
20 mA setting
a
)
Click OK button to enter 20mA parameter setting, crystal
shown as right.
b
)
Click OK button to enter 20mA value input interface, input
the value as right figure, indicated 20mA output when the level is 5m,
then click Mode to exit.
3
)
4mA setting
In the 20mA parameter setting menu click▼to enter 4mA
parameter setting menu, click OK button enter setting interface, method
as above.
4.4.3 Relay parameter setting
1
)
In a normal operation, click Mode button to enter
The parameter setting menu, then click▼to
choose P03(relay)relay setting menu, shown as right figure.
2
)
Click OK button to enter relay 1 setting menu, shown as right