11. TESTING HbA1c
The timer will countdown from 5 minutes until the HbA1c% is displayed.
08/03/30
HbA1c
AM10:08
Glucose
ID:
5.4
%
STEP 8.
After the test is done, open the cover of the analyzer. The display will show the 'cartridge
remove' icon. Take out the cartridge from the analyzer by gently pushing to the left and
pulling out.
STEP 9.
CAUTION :
Do not give strong impact to cartridge. Dispose all waste in
accordance with applicable national and /or local regulations.
NOTE :
If you see “>14.0%” or “<4.0%” displayed, HbA1c% level may be above
14.0% or below 4.0%, repeat testing. If the second result is again outside
this range, please contact your local customer support and service center.
08/03/30
HbA1c
AM10:09
Glucose
Remove
CLOVER A1c
™ Self
• 23 •
CLOVER A1c Self
• 22 •
11. TESTING HbA1c
Insert the reagent pack into the cartridge in the analyzer. The ‘close the cover’ icon will
be displayed.
STEP 6.
08/03/30
HbA1c
AM10:03
Glucose
The test starts automatically once the cover is closed.
08/03/30
HbA1c
AM10:03
Glucose
4:30
STEP 7.
NOTE :
Do not force the reagent pack into the system. The reagent pack is
designed to fit only in one way.
IMPORTANT :
Do not open the analyzer during the test.
Do not vibrate or move the analyzer during the test.