• 18 •
CAUTION
1. Control solution is for in-vitro diagnostic use only. This means that it is only used for
testing outside the body.
2. Control solution is exclusively made for the Finetest™ Lite Blood Glucose Monitoring
System. Do not use any other brand of control solution, made by other manufacturers.
3. Check the expiration date on the container. Do not use, if expired.
4. Once opened, use it for the next three months only. Record the discard date on the
control solution container. Discard after three months.
5. Do not swallow. This is NOT for human or animal consumption.
6. The control solution bottle icon contains letter ‘C,’ so as not to be confused with the
medication container. This will allow you to understand differences between a control
solution test and an actual blood test.
For example only
If you are using the normal control solution, according to this particular container of strips,
your meter should show a number between 4.3 and 6.7mmol/l.
Please note that the control values shown in the picture are not control ranges.
NOTE
- It is recommended that glucose control solution is stored at room temperature 20~25
℃
(68~77
℉
) before testing.
- Check the expiration date before performing a glucose control solution test.
- Do not use beyond expiration date. Once opened, the control solution expires after 3 months.
If your control solution test falls out of range, please follow the next steps before
contacting customer support:
- Check the expiration dates on all the products you are using. Be sure you are using
Finetest™ Lite Control Solution.
- Try another control solution test.
If this test falls out of range, try another control solution test with a new unopened
container of test strips.
If you continue to get control soulution test results that fall outside the range, do not
perform a glucose test. please call Neon Diagnostics Ltd.
H.
N.
L.
14.2-19.1
4.3-6.7
1.9-3.6
4.Dip the test strip into the control
solution liquid.
Results will appear in 5 seconds.
Compare the result to the range
printed on the test strip container.
Results should fall within that range.
3.Discard the 1st drop of
control solution. This will
eliminate any residue.
Place a drop of control
solution liquid on a clean,
dry surface (e.g. the lid of
the test strip container)
2.Gently shake the control
solution before use.
1.Check the expiration date before
performing a control solution test. Do
not use if expired. Please note the
expiration date is marked on the control
solution container. Once opened, the
control solution expires after 3 months.
Summary of Contents for Finetest Lite