INGOS Ltd. COR400                                                          17.7.2019

 

Strana 2  

 

1. 

INTRODUCTION 

 

 

The Chiller is designed to cool the cooler rotary vacuum evaporator (it can also be used for  

two rotary evaporators together). The chiller can be used for other device which need cooling. The 
control  module  controls  temperature  in  the  cooling  tanks.  The  chiller  is  designed  as  a  single 
functional instrument.  

                   4                                                 

 

3                                                                                                                            

                   5     

 

 

 

 

 

                   6 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

 

Switch 

    

 4.  Output cooling medium   

7.   Grip 

2.  Thermostat 

 

 5.  Input cooling medium 

 

8.   Socket or switch 

             3.  Circulation pump 

 6.  Overflow                                            circulation pump 

Fig.1   Chiller 

Warning 

Using devices other than as described in the instructions can cause security breaches! 

2. 

DESCRIPTION AND SPECIFICATION 

2.1  Description 

 

 

The instrument consists of a box in which is thermostatic tank, as a source of cold a modern 

compressor  cooling  unit  Lindr  is  used  that  uses  the  input  electric  energy  for  direct  transfer  to 
chilling.  Cooling loop in a thermostated tank is made of stainless steel.  Cooled thermostated tank is 
equipped with a circulation pump. Temperature control and cooler run is through digital thermostat. 

Used  materials  meet  the  highest  requirements  for  hygiene  standards  and  guarantee  a  long 

working life of this cooler.  

Two grips located on the sides of the cooler facilitate handling. 

 

 

 

 

Summary of Contents for COR 400

Page 1: ...rvice INGOS Ltd K Nouzovu 2090 143 00 Prague 4 Czech Republic Tel 420 225 983 400 420 225 983 409 Fax 420 244 403 051 e mail instruments ingos cz www ingos cz 17 7 2019 CIRCULATION CHILLER COR 400 Use...

Page 2: ...Chiller Warning Using devices other than as described in the instructions can cause security breaches 2 DESCRIPTION AND SPECIFICATION 2 1 Description The instrument consists of a box in which is therm...

Page 3: ...ct all packing material If any item is missing contact the manufacturer or your supplier Place the instrument on the table and familiarize yourself with the distribution and function of the controls a...

Page 4: ...e coupling it is necessary to moisten the end 3 2 Temperature controller Display along with a button keyboard provide user comfort for direct control The function of the individual keys is as follows...

Page 5: ...or in 30 seconds the thermostat when there is no button operation will return to normal operating mode Setting item Range setting parameters Factory setting code Temperature difference 1 C 16 C 1 C F0...

Page 6: ...f paint on the instrument The surface of the instrument clean with a soft cloth Cleaning the condenser perform at least 1x per month by sweeping or air pressure or CO2 Fan cover by pulling upward pull...

Page 7: ...be performed by the manufacturer or its authorized body Warning For using water as coolant set in mode F2 lower limit of the temperature at 1 C and above 7 WASTE DISPOSAL At the end of life discard th...

Reviews: