13
First use
1. After unpacking the system, check the device and all available
accessories for damage or defects, for example, device plastic body cracks
which may expose electrical components. Contact Innovatione Technical
Service or your supplier to report any damaged component.
2. Prepare 3 liters of distilled water.
3. Unscrew the two covers marked with numbers 1 and 2.
4. Insert the funnel hose into hole 1.
5. Pour water into the funnel until it flows out of hole 2.
6. Screw the covers back on.
7. The water needs to be changed every month.
8. To drain the water, you will have to unscrew the two covers marked with
the numbers 2 and 3. Before that, place a suitable 3-liter container under
the holes.
9. When the device was stored for a long time, it is strongly recommended
to replace the water in it before use.
10. It is strictly prohibited to transport the device with water in the cooling
system.
11. Plug the power cord into the mains outlet and hold its other end in
your hand. Insert the supplied mains tester screwdriver into hole L. If the
screwdriver diode is lit, the power cord is connected correctly. Otherwise,
reverse the power plug in the mains outlet.
12. Plug the power cord into the device power socket.