16
Device cleaning and
maintenance
1. Before cleaning, make sure the power cord is disconnected from the
mains.
2. Wipe the device body with a clean, damp cloth at least once a month.
3. When needed, clean the device ventilation openings, including those
located on the lower part of the device to prevent overheating.
4. The applicator needs replacing when 1,500,000 pulses are reached.
5. The applicator must be disinfected with an antiseptic agent before and
after each procedure.
6. Replace the distilled water in the cooling system once a month. See the
“First Use”
section to learn more about the replacement procedure.
7. Once in three months, pour in a proprietary cleaning liquid for
10 days
instead of distilled water.
8. Repairs, performed by unauthorized personnel will void the warranty.
9. In the full range of issues, contact your authorized supplier or Innovatione
Service Department.