background image

19 

 

12.

 

High Speed Handpiece 

 Maintenance and Care

 

In order to always provide the handpiece with clean air, water accumulated in the compressor must be drained out 

ONCE A DAY. DAILY LUBRICATION IS ABSOLUTELY ESSENTIAL.

 

12.1

 

Operation and Maintenance 

Operate the handpiece at a pressure of 35 PSI. The handpiece is engineered to attain speeds of 400,000 
RPM at 35 PSI. Use the brush to remove foreign particles. Particles can be dislodged by blowing air 
backward from the contra-angle head. 

NOTE: DO NOT

 attempt to blow particles from the rear end of the 

handpiece, as larger particles will be blocked in the water tube. 

12.2

 

AUTO-CHUCK (PUSH BUTTON) SYSTEM - INSTALLING AND REMOVING ROTARY CUTTING INSTRUMENT 

Grasp handpiece firmly so that your thumb can touch the cap, then push on centre of cap. Insert cutting 
instrument into chuck until it stops, then release thumb. NOTE: After installation, it is recommended as a 
safety measure to push button again one time (holding bur) and release. This will effect proper alignment 
of the jaws. For removing, grasp handpiece firmly so that your thumb can touch the cap, and push on 
centre of cap. Then, pull bur straight out. 

12.3

 

OBSERVE THE FOLLOWING SAFETY RULES 

Do not attempt to extend burs from the chuck. Always tug on the cutting instrument after chuck is loaded 
to confirm the cutting instrument is held securely. Before use, verify rotating concentricity by operating the 
handpiece outside the mouth in a safe place. Do not use instruments which do not run concentric or 
produce excessive noise. 

CAUTION:

 Never depress the auto-chuck end cap button or rest the end cap against any part of the oral 

anatomy during the handpiece operation. Depressing the end cap while bur is rotating can activate the 
chuck opening spring mechanism and result in end cap heat-up and/or loss of bur retention. 

12.4

 

Replacement of Cartridge (Turbine) 

a.

 

Place bur in cartridge. 

b.

 

Unscrew head cap with head button wrench, turning counter-clockwise. 

c.

 

Push bur on flat surface until cartridge is removed. 

d.

 

Clean inside of handpiece head to remove moisture and foreign particles. 

a.

 

Insert new cartridge into handpiece head. When replacing with canister style cartridge, the guide 
projection on the canister must be lined up with concave groove inside handpiece head to insure 
proper fit. 

e.

 

Replace head cap into handpiece by screwing clockwise with cap wrench. 

NOTE: When replacing the head cap, be careful to align threads correctly. these threads are very fine and 
can be easily stripped. 

12.5

 

Autoclave Procedure for High Speed Handpiece 

Remove bur from handpiece and clean handpiece with soap and water by scrubbing with a toothbrush or a 

2” x 2” piece of gauze, then towel dry the handpiece. Lubricate the handpiece. Insert bur back into the 

handpiece and connect to unit. Run the handpiece for approximately 15 - 20 seconds or until lubricant is 
out of handpiece. Remove bur from handpiece and disconnect handpiece from unit. The handpiece is now 

ready for sterilization. Follow manufacturer’s instructions for the autoclave unit being used. After 

autoclaving is complete, allow handpiece to return to room temperature, and then lubricate handpiece 
again prior to use. 

Summary of Contents for Veterinary Dental Cart

Page 1: ...1 F FI IB BE ER R O OP PT TI IC C H HI IG GH H S SP PE EE ED D V VE ET TE ER RI IN NA AR RY Y D DE EN NT TA AL L C CA AR RT T INSTRUCTION MANUAL Inovadent 200 Main Street Elmwood WI 54740...

Page 2: ...KLAW Tip Installation 11 5 2 KLAW Tip Technique 11 5 3 Power Settings Use of KLAW Tips 11 5 4 KLAW Tip Wear Guide 12 5 5 KLAW Tip Care Maintenance 12 5 6 KLAW Tip Angles 12 6 End of Day Procedure 13 7...

Page 3: ...cluding medical devices may interfere with normal operations of pacemakers It is suggested that patients or professionals who have pacemakers avoid using a Piezoelectric scaler The Inovadent Veterinar...

Page 4: ...4 2 High Speed Veterinary Dental Cart...

Page 5: ...ith distilled water and attach under left side of cabinet See diagram on Page 4 Item 5 6 Compressor is filled with oil at factory Check bubble window to insure oil level falls in between the MAX and M...

Page 6: ...se the pressure and counter clockwise to decrease the pressure Once the proper pressure is set push down on the adjustment knob to lock it into place 3 Verify the water system air pressure from 25 30...

Page 7: ...djustment will cause a significant change Follow the same procedure as above to adjust the Low Speed Handpiece the middle adjustment knob to 25 PSI and to adjust the Piezoelectric Handpiece the far ri...

Page 8: ...pray cleaner and lubricant such as Bio Lube is recommended If using simple oil as a lubricant place four drops into appropriate air Failure to oil after each use will void warranty Note Handpieces sho...

Page 9: ...e holder set the power dial and water flow and press on the foot pedal Water should be adjusted to a fine mist See Figure 4 8 and 4 9 Figure 4 8 Figure 4 9 The built in Piezoelectric scaling system po...

Page 10: ...not immerse in any solution including ultrasonic or cold sterile solutions as this will damage the handpiece and void Warranty Dry handpiece with a clean cloth or gauze square Place handpiece in a ste...

Page 11: ...eeded 2 2 3 Delicate finishing tip excellent for cats and small dogs Left 30 angle for difficult to reach surfaces Recommended for R mandibular buccal surfaces L mandibular lingual surfaces L maxillar...

Page 12: ...tion Care should be taken to prevent the tips from touching one another while autoclaved Heat from the autoclave cycle can deform or break tips if they touch one another Autoclave the tips at a maximu...

Page 13: ...AIR PURGE See Figure 6 3 Remove the High Speed Handpiece from the holder Hold the handpiece over a sink or a bucket and depress the foot pedal until water no longer sprays from the handpiece Remove A...

Page 14: ...he system using water 9 Remove KLAW tip from Piezoelectric Handpiece carefully clean and place in storage box 10 Remove bur from High Speed Handpiece and replace with blank bur 11 Store handpieces in...

Page 15: ...ocess See Figure 6 9 7 JUN AIR Compressor Review the JUN AIR Compressor Operation Manual before use DO NOT operate the compressor if the oil level is below the minimum line on the glass bubble window...

Page 16: ...les drain and dry Once dry wrap in bubble wrap and place inside cabinet 3 Remove air compressor intake filter and install small yellow plastic cap in its place 4 Coil all handpiece cords foot pedal co...

Page 17: ...instructions described in the instruction manual Merchandise returned as defective will not be replaced until it has been determined to be defective by Inovadent Please allow up to 3 4 weeks for this...

Page 18: ...te the installation Note This cannot be performed with a handpiece installed 3 Insert the bulb into the adapter body as shown MAKE SURE TO ALIGN THE BULB PINS WITH THE HOLES IN THE ADAPTER see Figure...

Page 19: ...ench turn the plug counter clockwise to loosen the plug See Figure 11 5 2 Add oil to the compressor through the plug opening To prevent overfilling add a small amount of oil at a time See Figure 11 6...

Page 20: ...m the micro mist filter by turning the filter clockwise Replace the old filter by turning the new filter counter clockwise Reattach the separator bowl by turning the bowl clockwise until snug See Figu...

Page 21: ...by pulling it apart Do not pull on the yellow wires Figure 11 13 Figure 11 13 3 Check the fuse for continuity If the fuse needs to be changed use the spare 1 Amp fuse provided with your unit Fuses ca...

Page 22: ...ot run concentric or produce excessive noise CAUTION Never depress the auto chuck end cap button or rest the end cap against any part of the oral anatomy during the handpiece operation Depressing the...

Page 23: ...rbine be sure to select one with the right rotation speed allowance Burs with a wider blade such as carbide bur tend to come off easily when cutting load exceeds chuck s holding force Bur also may com...

Page 24: ...utoclave at 135 C maximum Inovadent recommends Visually checking out your Inovadent instruments for any loosening and or damage immediately after each use Using protective clothing when handling Inova...

Page 25: ...Handpiece Reattach the handpiece to the unit and run the handpiece for 8 to 10 seconds Hold handpiece firmly Dropping the handpiece may cause damage b Surface Cleaning Clean with brush and wipe off by...

Page 26: ...o not run handpiece without locking the bur d Do not run handpiece with chuck opened e For safety mount test the bur the standard accessory on the Low Speed Handpiece when it is not in use f In using...

Page 27: ...LAW 3 Item 54830 Item 54831 Item 54832 KLAW 4 KLAW 6 KLAW E Tip 30mm Item 54833 Item 54834 Item 54840 KLAW E Tip 40mm KLAW E Tip 60mm KLAW E Surgical Item 54841 Item 54842 Item 54843 KLAW Wrench Piezo...

Page 28: ...iber Optic Item 54900 Item 54901 Fiber Optic Light Bulb 4 Hole Gasket 5 Hole Gasket Item 54905 Item 54911 Item 54910 High Speed Turbine End Cap for H S Handpiece Item 54903 Item 54907 Prophy Angle Pro...

Page 29: ...rd Length 330 Pear Shaped Bur Item 54820 Standard Length Item 54822 Surgical Length Item 54805 Surgical Length 699 Carbide Bur 700 Carbide Bur 702L Carbide Bur Item 54810 Item 54811 Item 54812 Standar...

Page 30: ...Parts Water Bottle Gasket Regulator Filter Micro mist Separator Filter Item 54912 Item 54918 Item 54916 JUN AIR Compressor Oil Water Bottle 1 Liter Ultra Scale Item 54884 Item 54925 Item 54270 Manual...

Reviews: