Inscale Ltd. 2010
p a g e
9
•
The display will show
“100.0
%
”
.
•
When another item with a different weight value is put on the pan, the
relative % value would be shown against the initial reference sample
value.
•
Press the
[Zero/
ESC
]
key to exit and go back to weighing mode.
5.3.3 CHECK WEIGHING
•
When in normal mode, press
[Func/
]
three times to move to the
check weighing mode.
•
The display will show
“CHECK”
.
•
Press
[Tare/
↵
↵
↵
↵
]
to set the high limits. The display will show the
“000000
HI”
with one digit flashing.
•
Press
[Hold/
]
to increase the flashing digits;
•
Press
[Units/
]
to decrease the flashing digits;
•
Press
[Func/
]
to select the digit;
•
Press
[Tare/
↵
↵
↵
↵
]
to accept the value and move to set the low limits. The
display will show
“000000 Lo”
•
Use the same method the set the low limits and press
[Tare/
↵
↵
↵
↵
] to
enter
the buzzer setting section,
•
Press
[Hold/
]
to change the display to
“H-L”
or
“- - 0- - ”
or
“OFF”
to
have beeper sound when the weight is outside the limits, between the
limits or off.
•
Press
[Tare/
↵
↵
↵
↵
]
to start the check weighing.
•
Put the item on the pan, if the weight is below the low limits, the scale
will show
“xxxxx.x Lo”
; if it is higher than high limits, the scale will
show
“xxxxx.x HI”
. Otherwise the scale will show
“xxxxx.x OK”
。
•
Press the
[Tare/
↵
↵
↵
↵
]
key to exit and go back to weighing mode.