NS-GLSBFSCW1
www.insigniaproducts.com
26
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment
IMMUNITY test
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD) IEC
61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC
61000-4-4
Not applicable
Not applicable
Surge IEC 61000-4-5
Not applicable
Not applicable
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC
61000-4-11
Not applicable
Not applicable
Power frequency
magnetic field IEC 61000-4-8
30 A/m
50 Hz/60 Hz
30 A/m
50 Hz/60 Hz
Conducted RF IEC 61000-4-6
Not applicable
Not applicable
Radiated RF IEC 61000-4-3
10 V/m
80 MHz - 2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz - 2.7 GHz
80% AM at 1 kHz
NOTE: U
T
is the AC voltage prior to application of the test level.