Maintenance
BlastMate III Operator Manual
6–1
6. M
AINTENANCE
This chapter outlines maintenance procedures for the BlastMate III including battery charging,
annual sensor calibration, cleaning, loading paper into the printer, and paper handling precautions.
6.1. Battery Charging
The BlastMate III internal battery requires periodic and regular charging. To check the battery
charge, turn the BlastMate III on and read the battery level from the Main Window. If a number
appears in this window instead of the bars, charge the monitor immediately to avoid possible
battery damage. Keep the unit on charge when not in use, or charge and then disconnect the
battery if storing for prolonged periods of time.
BlastMate III Main Window
Ready to Monitor
Battery: E - F
Memory: E - - - F
July 1 95 10:30:17
Ready to Monitor
Battery: E - - - - - - - - - - F
Memory: E - - - F
July 1 95 10:30:17
Low Battery
Fully Charged Battery
Figure 6.1 Battery Level Indicator.
WARNING:
Open the monitor’s lid to ensure adequate battery ventilation during charging.
Always charge the monitor in a well ventilated area.
Do not allow the battery to become discharged and left in a discharged state for
any extended period of time. Allowing the battery to become discharged may
cause permanent damage to the battery.
To charge the battery, plug the AC adapter into the connector marked PWR located under the
handle of the BlastMate III. Plug the other end into a 120 VAC 60 Hz electrical outlet. The battery
charging indicator light turns on. For electrical outlets other than 120 VAC 60 Hz, you must use
an optional power converter; see your dealer for more information.
All BlastMate III monitors use the DV-1250 AC adapter, provided with each monitor. The adapter
restricts the voltage and current to no more than 12 VDC at 500 mA. Voltage can increase to 13.8
VDC without damaging the monitor. For example, an AC adapter with an output rating of 12
VDC at 1000 mA may be used as long as the voltage does not exceed 13.8 VDC for loads over
500 mA when used with a monitor. Check the manufacturer’s specifications carefully to ensure
the AC adapter does not exceed the maximum voltage and current specifications.
Use of AC adapters that exceed the maximum voltage and current specifications may cause undue
stress on the monitor’s internal regulators. Like all electronic devices, exceeding these
specifications increases the risk of damage and possibly of an accidental fire. Instantel’s warranty
does not cover this type of damage.
Instantel recommends using the DV-1250 AC adapter supplied with each monitor. Replacements
can be ordered from Instantel. A 220 VAC adapter (p/n TC50A) is also available from Instantel
for use with the DV-1250 110 VAC adapter.
Summary of Contents for Blastmate III
Page 1: ......
Page 6: ...iv BlastMate III Operator Manual Notes ...
Page 20: ...Chapter 2 2 6 BlastMate III Operator Manual Notes ...
Page 50: ......
Page 74: ...Chapter 6 6 6 BlastMate III Operator Manual Notes ...
Page 77: ...Appendix BlastMate III Operator Manual A 1 Appendix ...
Page 78: ...Appendix A 2 BlastMate III Operator Manual ...
Page 84: ...Appendix A 8 BlastMate III Operator Manual Notes ...
Page 90: ...Index I 6 BlastMate III Operator Manual Notes ...