23
EN – ENGLISH
NOTE:
If additional extension height is desired the two stop knobs (# 8) can be backed out by
turning counterclockwise, thus permitting the extension arms to rotate higher.
5. Lock the slide lock knob (# 7) by turning it clockwise until the teeth have engaged on the
crossrail bracket assembly.
6. Lock the clamp bracket knob (#9) by turning it clockwise.
7. Lock the two extension knobs (# 5) and side lock knobs (#6) by turning them clockwise.
CAUTION!
CAUTION: Always be sure all joints and locking mechanisms are secure after
completing adjustments. Failure to do so may result in serious patient injury.
NOTE:
The extensions of the adaptor are stopped from passing below horizontal to prevent
potential collisions with the table during scanning. If necessary it is possible to override the
stop by turning the locking knobs #6 counter clockwise. The extensions can not be locked in
a position below horizontal.
Figure 24 Mounting Components