IntelliSAW IRM-48 Reader
Installation Manual
Page 22 of 28
IRM-48 Reader
910.00371.0001
6.2
M
ODBUS
R
EGISTERS
Note: table items highlighted in
blue
are not yet implemented.
REGISTER DESCRIPTION
TYPE
MIN
MAX
SCALE
ERROR CODE
418
1
temp 1
signed 16
-500
1675 0.1 C
0x8000
419
temp 2
signed 16
-500
1675 0.1 C
0x8000
420
temp 3
signed 16
-500
1675 0.1 C
0x8000
421
temp 4
signed 16
-500
1675 0.1 C
0x8000
422
temp 5
signed 16
-500
1675 0.1 C
0x8000
423
temp 6
signed 16
-500
1675 0.1 C
0x8000
424
temp 7
signed 16
-500
1675 0.1 C
0x8000
425
temp 8
signed 16
-500
1675 0.1 C
0x8000
426
temp 9
signed 16
-500
1675 0.1 C
0x8000
427
temp 10
signed 16
-500
1675 0.1 C
0x8000
428
temp 11
signed 16
-500
1675 0.1 C
0x8000
429
temp 12
signed 16
-500
1675 0.1 C
0x8000
430
ambient 1
signed 16
-400
1250 0.1 C
0x8000
431
RH 1
unsigned 16
2
0
1000 0.1% RH
0xFFFF
432
ambient 2
signed 16
-400
1250 0.1 C
0x8000
433
RH 2
unsigned 16
0
1000 0.1% RH
0xFFFF
434
ambient 3
signed 16
-400
1250 0.1 C
0x8000
435
RH 3
unsigned 16
0
1000 0.1% RH
0xFFFF
436
ambient 4
signed 16
-400
1250 0.1 C
0x8000
437
RH 4
unsigned 16
0
1000 0.1% RH
0xFFFF
438
ambient 5
signed 16
-400
1250 0.1 C
0x8000
439
RH 5
unsigned 16
0
1000 0.1% RH
0xFFFF
440
ambient 6
signed 16
-400
1250 0.1 C
0x8000
441
RH 6
unsigned 16
0
1000 0.1% RH
0xFFFF
442
ambient 7
signed 16
-400
1250 0.1 C
0x8000
443
RH 7
unsigned 16
0
1000 0.1% RH
0xFFFF
444
ambient 8
signed 16
-400
1250 0.1 C
0x8000
445
RH 8
unsigned 16
0
1000 0.1% RH
0xFFFF
446
SD1 Total
unsigned 16
0
65534 10 "Q
uhf
"
3
0xFFFF
447
SD2 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
448
SD3 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
449
SD4 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
450
PD1 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
451
PD2 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
452
PD3 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
453
PD4 Total
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
454
Data version = 1
unsigned 16
0
65534 na
0xFFFF
455
noise 1
unsigned 16
0
65534 10 "Q
uhf
"
0xFFFF
1
Registers start at 0. 418 is 400418 or 0x01A2 in the datagram.
2
RH may be treated as signed 16 making the error code translate to a small negative number.
3
Partial discharge and surface discharge are nonlinear, and the scale is approximate.