InTemp CX400 Series Temperature Logger Manual 

1-800-LOGGERS 4 

www.onsetcomp.com 

App only users:

 The logger includes preset configurations. 

To set up a custom configuration, tap the Settings icon and 
tap CX400 Logger. Also, if you need to perform daily logger 
checks, tap Record CX400 Logger Checks under Settings and 
select Once Daily or Twice Daily. This should be done before 
configuring the logger. 

See www.intempconnect.com/help for details on setting up 
custom configurations in both the app and InTempConnect 
and on setting up trip information. 

a.

 

Tap the Devices icon in the app. Find the logger in the list 
and tap it to connect to it. 

 

If you are having trouble connecting: 

 

Make sure the logger is within range of your mobile 
device. The range for successful wireless communication 
is approximately 30.5 m (100 ft) with full line-of-sight. 

 

If your device can connect to the logger intermittently 
or loses its connection, move closer to the logger, within 
sight if possible. 

 

Change the orientation of your phone or tablet to 
ensure the antenna in your device is pointed toward the 
logger. Obstacles between the antenna in the device 
and the logger may result in intermittent connections. 

 

If the logger appears in the list, but you cannot connect 
to it, close the app, power down the mobile device, and 
then turn it back on. This forces the previous Bluetooth 
connection to close. 

b.

 

Once connected, tap Configure. Swipe left and right to 
select a logger configuration. Type a deployment name. 
Tap Start to load the selected configuration to the logger. 

InTempConnect users:

 If trip information fields were set 

up, you will be prompted to enter additional information. 
Tap Start in the upper right corner when done. 

5.

 

Deploy and start the logger

. Deploy the logger to the 

location where you will be monitoring the temperature. 
Logging will begin based on the settings in the configuration 
selected. If the logger was configured to perform daily 
checks, connect to the logger and tap Perform (Morning, 
Afternoon, or Daily) Check every day.

 

Once logging begins, the logger will display the current 
temperature reading and the minimum and maximum readings 
within the current 24-hour period (midnight from one day to 
midnight the next day) if the logger was set to record logger 
checks (see 

Performing Logger Checks

). Otherwise the 

minimum and maximum readings represent the entire logging 
period and only reset when the logger is downloaded and 
restarted or stopped and reconfigured (they will not reset if you 
download the logger and continue logging). These minimum 
and maximum values are also available in the logger reports 
(see 

Downloading the Loggers

). 

For CX402 models, you can click Next to switch between 
external probe and ambient temperature readings. Minimum 

and maximum readings are only available for the external 
probe. 

Logger Alarms 

There are two conditions that will trip an alarm: 

 

When the temperature reading by the external probe (if 
applicable) or the ambient temperature is outside the 
range specified in the logger configuration. 

 

When an external probe (if applicable) is disconnected 
during logging. 

When a temperature alarm trips: 

 

The logger LED will blink every 5 seconds. 

 

The alarm icon will appear on the LCD and in the app. 

 

The logger will beep every 15 seconds (unless audible 
alarms are disabled in the logger configuration). 

 

An Alarm Tripped event is logged. 

When an external probe is removed: 

 

The logger LED will blink every 5 seconds. 

 

“ERROR” and “PROBE” will appear on the LCD and 
“ERROR” will appear in the app. 

 

An alarm icon will appear in the app. 

 

The logger will beep every 15 seconds. 

 

A Probe Disconnected event is logged. 

To mute a beeping alarm, press the Mute button on the logger. 
Once muted, you cannot turn the beeping back on. 

Download the logger to view details about a tripped alarm and 
to clear alarm indicators in the app and on the LCD (the probe 
must be reconnected for ERROR to clear on the LCD). For a 
temperature alarm, the tripped alarm will clear once the logger 
is downloaded and restarted. 

Passkey Protection 

The logger is protected by an encrypted passkey automatically 
generated by the InTemp app for InTempConnect users and 
optionally available for local users who are using the app only. 
The passkey uses a proprietary encryption algorithm that 
changes with every connection. 

InTempConnect Users 

Only InTempConnect users belonging to the same 
InTempConnect account can connect to a logger once it is 
configured. When an InTempConnect user first configures a 
logger, it is locked with an encrypted passkey that is 
automatically generated by the InTemp app. After the logger is 
configured, only active users associated with that account will 
be able to connect to it. If a user belongs to a different account, 
that user will not be able to connect to the logger with the app, 
which will display an invalid passkey message. Administrators 
or users with the required privileges can also view the passkey 
from the device deployment page in InTempConnect and share 
them if needed. See www.intempconnect.com/help for more 
details. 

Local Users 

If you are not using InTempConnect and instead logging into 
the app as a local user, you can create an encrypted passkey for 

Summary of Contents for CX402-T215

Page 1: ...rift 0 1 C 0 18 F per year Response Time Probe only 1 minute 30 seconds typical to 90 in air moving 1 m s With 5 mL bottle 10 minutes typical to 90 in air moving 1 m s With 15 mL bottle 20 minutes typ...

Page 2: ...ambient internal sensor temperature readings Magnets There are four magnets on the back of the logger not shown in diagram for mounting Temperature Sensor This internal sensor measures ambient temper...

Page 3: ...s been downloaded the LCD will turn off automatically after 2 hours The LCD will turn back on the next time the logger connects with your device Getting Started InTempConnect is web based software whe...

Page 4: ...download the logger and continue logging These minimum and maximum values are also available in the logger reports see Downloading the Loggers For CX402 models you can click Next to switch between ex...

Page 5: ...ed formats for sharing at a later time Tap the Reports icon to access previously downloaded reports See www intempconnect com help for details on working with reports in both the app and InTempConnect...

Page 6: ...condensation If the logger gets wet remove the battery immediately and dry the circuit board Note Static electricity may cause the logger to stop logging The logger has been tested to 8 KV but avoid e...

Page 7: ...one of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from th...

Reviews: